Home Contact us About us Projects

OUR MAKER GROUP

© Our Maker Group 2020

Stylus Balance


The problem

I have been using a Technics SL-DL1 turntable for my record collection since they came out in 1983, and it has an automatic, dynamic pressure control for the stylus. This, therefore, did not need me to test the stylus pressure so no device was needed to measure it. It is still working as good as new and produces good quality hifi sounds.

But the time has come for an upgrade, and the turntables I am looking at need very careful manual adjustment of the stylus pressure and I need something to do it with.

There are a number of cheap electronic devices on the market (all under different brands, but obviously coming from the same Chinese factory) claiming to measure to 0.05g but they are suspected of being of variable quality, reliability and accuracy.

It is not uncommon in Chinese factories to have several production “lines” making the same-looking product, but costs are saved by using different sources or specifications of the internal components according to the sponsoring purchaser. This could explain why such a variety of seemingly identical products advertised on Ebay are offered at such different prices. The other problem with anything electronic that you don’t use very often is that the batteries have always gone flat whenever you need them.

There are good quality items available but they are seriously expensive. So I thought I would make my own. Couldn’t be very difficult, could it?

The specification

I wanted a device that could measure to an accuracy of 0.05g over a range of 0.7g to at least 1.5g, with an option to extend the range if required, perhaps for other purposes.



The design

I considered a number of options and decided on a balanced beam not unlike the simple see-saw of the AR device. But to use a set of weights capable of measuring 0.05g steps over the required range would be unmanageable. Also, to make small adjustments could involve moving a number of weights, and they would all need to be accurately calibrated.

I therefore decided to use the same principle as hospitals used to use to measure the standing weight of patients, but perhaps a bit smaller! The idea is to have some sort of balanced beam with a sliding balance weight that is moved along a calibrated scale, preferably notched to ease registration with the calibrated values. Obviously, both the scale and the balance weight need to be accurately calibrated, and the scale needs to be easily read to the required precision.

I wanted the most likely range of values to be within the scope of a single balance weight along a shortish beam, but with the option of using a different balance weight to extend the range if necessary.

The materials

The beam itself is simply a gash length of 3mm stainless steel wire cut from a builders’ wall tie. It could have been any steel wire salvaged from some defunct bit of kit, but preferably not much thinner.

The beam head is a T-shaped block and drilled in three places. The horizontal hole carries the beam centrally, and the two vertical holes are for the pivots.

The pivots themselves are 3mm screws or bolts bevelled to a point, and long enough to project at least 2mm through the T-block.

The pivot screws are held in place with hex nuts recessed into the beam head.

The pivots rest on a U-block to hold the beam high enough to swing by at least 10 degrees each way.



The balance weight is a piece of copper wire salvaged from some 40A mains cable. Other wire could be used, but the reason for using electrical wire is that the specification can be looked up to find the weight per metre, so a calibrated weight can be made simply by accurately measuring the length.

The test weight is similarly cut at a different length.

The tooling

A vernier gauge was used to measure the length of the beam (90mm) and to find the exact centre datum.

To cut the stylus pit, and mark the calibrated balance beam notches a hobby drill with a drill stand and milling bed was used. This was convenient and accurate, but I expect there are many other ways of achieving the same thing.

The T-block and U-block were printed using a 3-D printer in PLA. They could have been cut from solid material (plastic or metal) but PLA is strong enough and a lot easier for me.

The calculations

The first thing was to find out how a simple wire beam could accurately register the point of a stylus at one end, and carry calibrated notches for a balance weight at the other.

The first is easy, as it comprised simply of filing a flat and drilling a pit at an exact distance (40mm) from the centre mark. To keep the beam symmetrical a corresponding flat was filed at the other end of the beam, but underneath so it didn’t get in the way of the calibrated notches.

The second is more involved as it means calculating the positions of the notches so that the steps correspond with simple values in stylus weight, e.g. 0.05g.

In basic terms we are looking for equal torque pulling downwards on both sides of the balance. This is made easier if the beam itself is balanced about it’s exact centre, which is why the two ends have exactly the same exposed length. The beam itself can then be disregarded for the remaining calculations.

The downward torque from  either the stylus or the balance weight is simply the product of weight and distance from the pivot (beam centre). So if the stylus is close to one end of the beam, and the balance weight is equal to the maximum measurable weight (1.5g in this case), the notch required to achieve balance at maximum stylus weight is simply the symmetrical position at the opposite end of the beam.

By moving the weight nearer to the centre, the torque correspondingly reduces and affords a balance with lower stylus pressure without needing a different balance weight or having to place the stylus anywhere different.

It was decided that 90mm beam length was a convenient size to place on a turntable deck. This allowed the stylus and final notch to be placed 40mm each end from the pivot, leaving 5mm at each end.

Having set the limiting notch, the distance between notches can simply be calculated from the balance weight (1.5g) and the required step in weight (0.05g).

Each measured step (0.05g) is 1/30 of the maximum measured weight, so the distance of each step is 1/30th of the distance to the maximum weight distance (40mm). This gives a figure of 40/30mm, which is 1.33mm. So the remaining notches are cut at intervals of 1.33mm.

Measuring to 1/100th of a millimetre is neither easy or necessary. The important thing is to ensure every third notch is at 2mm intervals, and the intermediate ones are spaced as nearly as possible in thirds. An error even as big as 0.25mm corresponds to a weight measurement of only about 0.01g, which is too small to worry about for the present purposes. However, all measurements should be taken from the end notch as datum so there are no cumulative errors, which could otherwise become significant.

To measure the required range of 1.5g down to 0.7g, a total of 16 steps of 0.05g are needed, requiring a total of 17 notches. As can be seen from the pictures, the design happily allows for a few further notches to be used, allowing for correspondingly lighter weights.

If heavier weights are to be measured, a heavier balance weight can be used. For example a balance weight of 3g can be used to measure 1.4g to 3g in 0.1g steps. This gives an overlap with the 1.5g weight.


Contributed by:

Peter Grossi

Construction

The beam is prepared by cutting to length and filing a flat for the stylus. This represents the top of the beam. A corresponding flat is filed underneath the opposite end to balance it out.

The stylus flat I marked with a drill to register the stylus accurately when used.

The calibration notches are cut into the top  of the beam at the other end.

The T-block needs to be a close fit for the beam.

The screws used as pivots need to be filed to points as centrally as possible. Fitting them into the chuck of a drill and grinding them on a fairly fine file should be a good substitute for a lathe. The finer the point the better. The screws need to be long enough to project at least 2mm through the T-block although they will probably only be required to project about 1mm.



Calibration

The beam is centred in the T-block, ensuring the distance of the stylus mark and the furthest notch are at exactly the same distance from the T-block.

With the screws in place and showing about 1mm of the points the beam assembly is placed on the U-block. The screws can then be adjusted to the T-block swings vertically in the centre. This can be done well enough by eye.

If the beam will not stay horizontal it may need balancing by carefully slipping it through he T-block one way or the other. If this requires the above distances to vary by more than 0.1mm, it may be necessary to file a tiny bit from one end of the beam. This may arise because of the weight lost in cutting the notches, or from the two flats not being exactly equal.

If the beam falls either way but won’t stay in the middle the sensitivity needs adjusting by retracting the two screws by the same amount.

When finished the beam should happily stay horizontal, and will oscillate slowly about the horizontal if provoked.

Ensure the stylus notch is facing upwards.

The beam needs to be sealed in place when calibrated. I used a drip of Humbrol enamel each side rather than superglue. Sealing is important to ensure it doesn’t move off centre when handled (or dropped!).


The balance weight

Clearly the accuracy of the balance weight is reflected directly in the measurements, so it needs to be as accurate as possible. Kitchen scales and postal scales are not nearly accurate enough, but there is an easy way.

Electrical wire (solid core) is specified by size and weight, which can be looked up. For example, the wire I used is taken from 40A house mains wiring and weighs 20g per metre. So for a 1.5g weight I needed 75mm. This can be measured with an ordinary ruler, as an error up to 1mm is well within the required tolerance, although you might prefer to use a vernier.

By folding the wire into a sort of pretzel it becomes easier to manage without dangling below the beam and grounding.

I cut a second weight using 50mm of wire to make 1.0g, and used this as a test “stylus” to check the beam calibration. It should be balanced by the balance weight at the 7th notch from the inside. Calculation: 1.0g – 0.7g = 0.3g; with 0.05g steps, this is 6 steps, which puts it on the 7th notch.

The picture below shows the whole assembly in balance, with the pretzel-shaped 1.5g balance weight on the 7th notch, and the test 1.0g weight draped over the stylus pit.


Usage and care

Before using the balance ensure the T-block is centred in the U-block, clear of rubbing against the sides.

Check the beam balances on it’s own without weights in case it has been damaged since last used, and that it oscillates freely when provoked. Then use the test weight if you want to be really sure.

When setting a stylus weight by this method it is usual to set the balance to the required weight, place the stylus on the balance , and then adjust the tonearm balance weight to achieve a level balance beam.

To measure the actual pressure of a stylus, place the balance weight to match the best guess of the expected weight, place the stylus and carefully slide the weight up and down to achieve a balance. Read the weight by counting notches (0.05g per notch, working down from 1.5g at the outer notch, or up from 0.7g at the inner notch)

Keep the balance beam, the U-block, the balance weight(s) and any test weights safely when not in use, although they are reasonably robust against accidental misuse.


Prior to the Technics I had an AR turntable, which required manual setting of the stylus pressure, and it came with a rather natty little plastic see-saw balance with balance weights that could measure in 0.25g steps to a maximum of 3.75g (2g, 1g, 0.5g and 0.25g). See pic. Unfortunately, the cartridges I have in mind require a pressure of about 1g (generally in the range 0.7g to 1.5g) and I would want to measure it to better than 0.1g if possible. So I needed something new.